La lectura de temperatura del sensor Z es unos grados más cálida o más fría de lo esperado.

Sensor Z.png

Check the sensor position

Check the sensor position if the sensor is near an exhaust fan or a cold air fan. Micro-climates can cause inaccurate readings. Our sensors are rated to be within ±0.5°C (±0.9°F) and can be NIST/ISO certified by an accredited lab to ensure this reading accuracy.

Compare to other sensors

Al comparar las lecturas de temperatura de nuestros sensores con las de otro termómetro o registrador de datos, es importante tener en cuenta lo siguiente:

  1. The margin of error of the comparison tool: Many thermostats will have a much higher margin of error than our sensors, ±0.5°C (±0.9°F).

  2. The temperature gradient of the room, refrigerator, or any other placement is in three dimensions: Measuring the temperature in one part of the location can yield different results than in another location due to micro-climates.

  3. La temperatura del aire cambia más rápidamente que la temperatura de una gran cantidad de líquido: La temperatura del aire en un frigorífico puede aumentar drásticamente cuando se abre la puerta, pero la temperatura del contenido del frigorífico aumentará más lentamente debido a su mayor masa térmica. Recomendamos el uso de un vial tampón (virtual o físico) para garantizar que se obtienen lecturas adecuadas en todo momento.

  4. Las pistolas de infrarrojos miden la temperatura de la superficie, mientras que nuestros sensores miden la temperatura del aire ambiente o las temperaturas amortiguadas.